Polypropylene plates are less adsorptive than polystyrene plates. However, they are not especially transparent so aren't much good for absorbance assays.
I find that the adsorption problem can be minimized in standard untreated polystyrene plates by including a bit of nonionic detergent, such as 0.01% Triton X-100.