I work with a Sciex Qtrap 4500 Hybrid Triple Quadrupole Linear Ion Trap mass spectrometer and am having difficulty observing compounds in EMS (enhanced full scan) mode.

Running a DI water blank results in the same mass spectra as running a 20mg/L sample of RDX. For both samples the TIC is always in the 10^6 - 10^7 range. This EMS method uses the almost all the same parameters as an MRM method that results in a well defined ~10^6 intensity peak at 1.7min for the most sensitive transition (266.8/92.1). However no 266.8 peak is present in the EMS method at the retention time or elsewhere throughout the scan. These scans are attached.

Previous tests of the EMS scan mode using a different method saw similarly high TIC with many peaks present in blank samples, including injection of water directly into the ion source by syringe, so the issue does not appear to be column bleed.

Parameters used are as follows:

Source Temp: 360C

Curtain Gas: 20

IonSpray Voltage: -4500

GS1: 60

GS2: 60

Solvent A: MilliQ Water + 10mM ammonium formate

Solvent B: UHPLC Grade Methanol + 10mM ammonium formate

Column: Hypersil Gold PFP 100x2.1 3um

MRM DP: 20

EMS DP: 30

EMS CAD gas: High

EMS CE: 10

Any suggestions as to why there is so much noise, and why the analyte of interest does not appear?

More Michael Knapp's questions See All
Similar questions and discussions