20 November 2020 5 5K Report

I'm running a Waters UPLC-TQD system. I've been experiencing challenges with repeatability of peak areas across analytes , columns and methods.

I'm currently trouble shooting the issue. I'm running repeat injections of a Caffeine standard. The concentration is 100 ppb, injection volume is 10 ul. Areas start off being repeatable, then drop and change . Minimum value = 5455, maximum = 14419, median 13081. Signal-to-noise ratio is also all over with a minimum = 40870, maximum = 585158 and median =161406.

I've checked that there aren't pressure drops / spikes on the UPLC, no leaks that I can detect (pressure test passed, but I can't test up to the pre-column heater as I don't have a blank to blank off at this point). I've repeatedly primed and purged the injector, the purge solvent is mobile phase.

On the MS side I've done a resolution tune (passed) and have calibrated.

Any comments / suggestions as to how I can improve repeatability? %RSD = 25

More Nial Harding's questions See All
Similar questions and discussions