I want to analyze lube oil concentration in water. Therefore I use a device working with fluorescence. The oil is extracted with hexane from an oil-water sample and the fluorescence of this solvent phase is measured. The device worked well with crude oil before. 

When changing the oil I obviously did a new calibration with the lube oil.

As proposed in the manual I prepared two standards (1. blank hexane, 2. 5000 ppm oil in hexane) and calibrated the device successfully. However, if I analyze a sample from the oily water (extraction with hexane), the values are much too high (100 times higher than expected). 

Next thing I did was trying to do the calibration with the solvent phase from an oil-water sample with hexane (oil+distilled water+hexane) were I knew the concentration (300 ppm). During calibration, the device showed "too high", although the oil concentraition is well below the maximum value.

Can it be, that the distilled water itself or any reaction products go to the solvent phase, fluoresce a lot and thus yield wrong values? Have anyone heard of this phanomenon?

More Johanna Ludwig's questions See All
Similar questions and discussions