After a satisfactory refinement by Fullprof I want to go for microstructural analysis. But I have no IRF file. How can I make a IRF file from my pcr file or how can I get them.
3) change the differents values of the parameters depending of the conditions acquisition (lambda, 2Theta-min, ...), and the refined parameters of your standart (U,V,W,X,Y,Z)
4) save the file and change the Filename extension as "irf"
Thank you Dr. Lefeve for your details elaboration.
I have no Idea about standart, i am afraid. I refined my samples with Pseudo-Voigt function. Please let me know what are required to do micro structure analysis using Fullprof.
Both diffractometer and sample contribute to the broadening of the peaks. In fact the observed profile is a convolution of the diffractometer contribution and the sample one. Therefore if you want to study micro-structure of the sample, you have to do correction of the instrumental contribution. that can be done using a standart (e.g. Al2O3, LaB6, ...) for which the refined parameters are encompassed in the irf file. Fullprof uses then this file to do the correction.
The IRF file contains parameters that account for the contribution of a diffractometer to the observed peak shape. Different instrument will contribute differently to the profile shape, hence it is recommended that you investigate the contribution of your instrument. You will have to find out the values for your instrument by running a full range for LaB6 (SRM 660). You can then deduce the IRF parameters by plotting FWHM^2 against Theta. From which you can use Caglioti's formula ( FWHM^2 = U tan^2 θ + V tan θ + W ) to deduce your IRF parameters - U, V and W.