We recommend using 50mM EDTA, followed by many rinses with pure water. These should be followed up with rinses of MeOH/Water (1:1). That usually removes most metal impurities. If the column is truly fouled, then replace it with a new one and be sure to use proper washing methods AFTER each analysis to prevent this.