When starting the instrument, the roughing pump would run normal and reach vacuum to 2 or 1 torr, the turbo pump would accelerate and reach normal status, however after a short time running at such normal pumping down process the ion gauge would just start momentarily to show a number around 20-40 and then immediately the turbo would shut down and then after a while the turbo would try to pump down again. It sounds like a vacuum leak but there is no way I can think of where that leak would be. Since the roughing vacuum seems OK, I replaced the ion gauge and one of the O-ring on the Q0, but that didn't solve the problem. What do you all think would be the problem and how to check it? Could the turbo be not up to speed? Something wrong with the ion gauge controller? Thank you all.