I have synthesized silver nanoparticles. The Sem image give 45 nm but TEM image give 4-5 nm particle size. Why there is big difference for the same nanoparticles?
SEM is used for imaging the surface of the nanoparticle in a submicroscopic scale . In contrast, use of TEM is to image the internal structure of the nanoparticle in a nano-meter scale.. simply TEM has a higher resolution than SEM by a factor of 10 or more. ... In a TEM, a nearly parallel beam of electrons travels through a thin specimen, and the resulting image is magnified electron-optically by a series of electromagnetic lenses, the main one of which is the objective lens. That why Particles appear ∼3% larger in the SEM than in the TEM microscope.
SEM may not be adequate to give the needed resolution to make assessments of nano-particle sizes. Thus, the particles are not well resolved and magnified. This fact makes it impossible to make accurate estimates of the particle size distribution of the SEM image of nano-samples.
Dear Mohammad is it possible that organic layer on synthesized nanoparticles is dissolved in solvent (DMF) during sample preparation for TEM. that's why it give very less size value than SEM
All images taken by TEM are artifacts of sample preparation (e.g. dispersion in epoxy and microtoming) and selection of particles. Consider microtoming a cube. See attached from 'A Primer of Higher Space' by Claude Bragdon The Manas Press Rochester and New York (1913)
In both cases (TEM & SEM) the imaging and calculation of 10000 particles would be needed to specify the mean to 1%. If the x90 is required then 10000 particles would be needed in the region above x90. Thus, in both cases it's highly unlikely that a representative sample is taken. You can measure particle size but not reliable particle size distribution with both EM techniques.
Last, but not least, the harder you look then the more you see. The same effect can be shown with optical microscope measurements where, when the objective lens is increased in magnification, then the measured sizes decrease. The comment by Madhukara Naik M is succinct and true.
Most likely a sampling issue. I think you can try embedding the nanoparticles in resin, obtain a thin section on a TEM grid and locate the same nanoparticles under SEM and TEM. Alternatively, you could simply disperse your nanoparticles in DI water and place a drop on carbon coated grid. This would be a direct size measurement comparison. If the difference is huge, then your technician needs to check the calibration.
Most general differences have been discussed in this thread: https://www.researchgate.net/post/What_is_the_difference_between_SEM_and_TEM_techniques2/1
So for nanoparticles, I'd like to point out that SEM has lower resolution (nm-range) and is able to scan larger areas than TEM. It's not suitable for very small particles (
Enoch Victor SW Wong You indicate precisely where TEM goes wrong:
A tiny amount of sample (< pg usually) is taken which is not representative of the whole
Enough particles are not measured for statistical validity (10000 random particles to get the standard error on the mean at 1%; tails of distributions need the same constraints)
The sample preparation stage produces a size and shape distribution even if there wasn't one to begin with. Look at the single cube above in the Blagden pictures. Consider slicing monodisperse spheres - a set of discs from effectively 0 up to the true diameter of the sphere. Average will be 1/2 the true value in this situation
I would only ever use TEM for lattice and atomic spacings in crystalline material. It is not capable of delivering a true size or shape distribution
SEM is used for imaging the surface of the nanoparticle in a submicroscopic scale . In contrast, use of TEM is to image the internal structure of the nanoparticle in a nano-meter scale.. simply TEM has a higher resolution than SEM by a factor of 10 or more. ... In a TEM, a nearly parallel beam of electrons travels through a thin specimen, and the resulting image is magnified electron-optically by a series of electromagnetic lenses, the main one of which is the objective lens. That why Particles appear ∼3% larger in the SEM than in the TEM microscope.
Most likely due to lower resolution in SEM analysis, you observe the paticles with two or three or ... number as an individual particle with larger size. So, when you are observing the microstructure via TEM, you behold the particles with smaller size!
I do not know which SEM microscope do you use and how conductive is the sample. Howewer, if both SEM and sample is ok, you shloud be able to see particles of diameter 4 nm... at least you can feel that they are there.