Our goal is to decrease the quantification limit as much as we can. We have a HClass-Xevo TQD Waters with a BEH C18 column 1.7 100 mm
simple scheme in our lab if we would like to decrese LOQ without much change in method:
extraction step:
increase extraction volume of plasma
increase volume of extraction solution or extract twice
use SPE
combination above methods
detection and quantification:
increase/decrease injection volume - LC step
increase/decrease injection volume - MS step (change split, change capillary distance etc)
other possible- change method at all - play with conditions, play with different m/z, different mobile phase, column . it is a huge field to play on.
SPE, as noted by Daria, or SPME. SPME will give you a significant concentrative effect.
How to calculate the RMSD values for a MD simulation using MOE?
07 August 2021 0 0 View
When I tried to energy minimization my system, I got fatal error as below. Fatal error: Atomtype opls_116 not found Although I've already added this line: ; include water #include "oplsaa.ff/spc.itp" to [molecultype] directive in my topology.
16 June 2021 0 0 View
effect of double bond on the flexibility of polymers
07 April 2021 0 0 View
I am writing to request some information for making polymer color flakes. I would appreciate the help.
17 March 2021 0 0 View
I took measurements of different concentrations of 15 nm and 5 nm Au nano particle solutions. I'm trying to determine which concentration I measured gives the best reading overall. I also...
02 March 2021 7,626 1 View
I am at the beginner level in drug design and drug discovery, and I am interested to learn more about this field. Which books are highly recommended to start from and gives me all essential...
02 March 2021 8,437 2 View