Our goal is to decrease the quantification limit as much as we can. We have a HClass-Xevo TQD Waters with a BEH C18 column 1.7 100 mm
simple scheme in our lab if we would like to decrese LOQ without much change in method:
extraction step:
increase extraction volume of plasma
increase volume of extraction solution or extract twice
use SPE
combination above methods
detection and quantification:
increase/decrease injection volume - LC step
increase/decrease injection volume - MS step (change split, change capillary distance etc)
other possible- change method at all - play with conditions, play with different m/z, different mobile phase, column . it is a huge field to play on.
SPE, as noted by Daria, or SPME. SPME will give you a significant concentrative effect.
Hello, Why do i see this baseline drift when i compare my blank (black) to the sample (blue)? Any suggestions as to why this happened? Thank you!
11 August 2024 3,770 4 View
I'm currently working on calculating the collision cross section (CCS) for various ions, and I'm facing challenges when dealing with sodiated and multiply charged ions. Most of the resources I’ve...
08 August 2024 8,329 0 View
I am analyzing cherts from different sources found within one specific geologic formation using ICP-MS/ES. My committee was asking this question about my analysis. I was not sure how to answer this
05 August 2024 7,902 3 View
Have you ever seen a LC-MS/MS method uses both internal standards and external standards (in matrix matching purpose) but the concentrations of internal standards are outside the calibration curve...
05 August 2024 3,084 6 View
I am currently working on a project involving liposomes and need to determine the maximum volume of siRNA that can be added to a 2.5 mL liposome solution with a total lipid concentration of 10...
30 July 2024 6,420 1 View
Can an analytical method's limit of quantification (LOQ) be outside its linear dynamic range, or is it always required to be within it? Please provide a thorough explanation supported by verified...
29 July 2024 7,198 9 View
I am doing research on wear mechanism of the additive manufactured metal parts under dry condition. Please share the reference to start the analytical modelling of the wear prediction.
29 July 2024 6,115 0 View
Dear Siesta Users, I installed Siesta 5.0.1 and I want to use Grimme's D3 correction. According to the 5.0.1 manual, I could use a few parameters to include the D3 correction in the...
27 July 2024 5,748 0 View
It has been long known that EDC carbodiimide and NHS easter's half life agains hydrolysis are highly related to pH. However, very surprisingly, not yet found a paper giving a table/chart of their...
25 July 2024 8,738 0 View
I am using the Polyomics Integrated Metabolomics Pipeline (PiMP) for metabolite feature analysis. After annotating my data, I have encountered cases where multiple potential metabolites are...
24 July 2024 5,789 0 View