Hello everyone
I'm doing evoked EPSC recording in 250µm hippocampal slice, I'm stimulating using a bipolar tungsten electrode from NPI (500µs)
the problem is that I can recruit EPSC but the amplitude will decrease gradually and rapidly so I can't have a stable baseline. I don't think it's a rundown problem since the problem to less degree persist in perforated patch clamp
I'm also recording spontaneous EPSC I can see a lot of them at the start of the exp but once I add GABA(A) and GABA(B) blockers most of those disappears. In my experiment, I'm recording isolated AMPA/NMDA currents
Any insight to help me troubleshoot the problem, I'd greatly appreciate it
My stimulator is an older model of this one https://www.npielectronic.com/product/iso-stim-ii/
and my stim electrode is a Cluster Eletrode, 2-contact IP=2x25um Pt/Ir from FHC positioned at the chaffer collateral close to the CA3 and with moderate distance from the recording electrode