Dear network,

Around six months ago we validated nitrosamines method on a Sciex +6500 / Agilent 1290 System and signals were great (Based to FDA method); after this numerous analysis we have done responded according validation.

Recently we are experiencing a contamination for NDBA signal (159 m/z --- 103 m/z) which is presented in every injection with a 2-3 x 10e6 magnitude.

We have tried the following actions in order to clear the contamination without a positive response:

Sciex +6500

-

Agilent UHPLC 1290

- AS injector valve cleaning

- AS module exchange with another UHPLC.

- Full UHPLC cleaning and washing for 14h with acetonitrile: IPA: cyclohexane: dicloromethane (25: 50: 15: 10)

- Colum washed for 7h.

- New column.

Others:

- new mobile phase and diluents preparation from unopened solvents.