I want to prepare ultra low concentration formaldehyde standard gas for sensor calibration, like 3~10 ppb concentration level. With no experience in gas filed, any suggestions or experience sharing are appreciated.
Gas standards in cylinders are not available at those low concentrations. You can get standards or even commercial mixtures as low as 2 ppm, so then you can perform a dilution step to the desired concentration. As formaldehyde is very reactive, dynamic standards (such as http://www.sciencedirect.com/science/journal/00399140/108) are preferred to static standards (gas cylinde rs). However these are pretty expensive to set up, so it all depends on what resources you htave and how accurate you need to be.