Dear All,
I have a pH electrode of BNC probe type ,which will give the output as mV .Can any one help me to convert sensor mV values into respective pH values .
Any suggestions /help is highly appreciated .
pH=7 - millivolt current=0
For growing ph, remove roughly 57.14 mV per unit of pH (ph=8, millivolt current=-57.14)
For curbing ph, add roughly 57.14 mV per unit of pH
If you are lazy, use (for example):
http://www.traditionaloven.com/conversions_of_measures/ph-voltage.html
Thanks Michel ,
I have interpreted like this pH = 7 - mV/ 57.14 .
I am afraid there might be a misunderstanding... I give you some examples:
pH5 --> mV: 114.29
pH6 --> mV: 57.14
pH7 --> mV: 0
pH8 --> mV: -57.14
ph9 --> mV: -114.29
And so on...
No Michel there is no misunderstanding ,I understood what you meant .
In my formula ,I can also able to calculate fractional values .
Formula --- pH = 7 - mV/ 57.14 .
For example when mV value = 57.14 i will get pH value of 6 by substituting mV value in the formula i .e pH = 7 - (57.14/57.14) = 7-1 = 6 .
lly for -57.14 pH = 7 -(-57.14/57.14) = 8
for -114.29 pH = 7 -(-114.29/57.14) = 9
This can also help to get fractional values of pH.
for 64.2 mV the pH value will be pH = 7 -(64.2/57.14) = 5.88
Kindly let me know if i am wrong .
1. Choose two buffers, let say pH1 = 7.00 and pH2 = 4.00 and measure the corresponding E1 and E2 (in mV).
2. For a sample with pHx, you measure Ex.
3. Calculate pHx = pH1 + (Ex – E1)(pH2 – pH1)/(E2-E1)
Important notes:
- all buffers and samples must be measured at the same temperature
- use a stabilisation time of about 1 minute for the electrode response
- the instrument must have a very high impedance (500 Gohm, not used in common voltmeters; use a pH/mV-meter in "mV" mode to test the equation)
- use a 3rd known buffer as sample (e.g., pH= 9.00 or 10.00) to verify if the sensor response is linear).
- at pH=7.00 , the common measured E is not zero; in the given equation, (Ex – E1) is the correction for this asymmetry
- in the given equation, (pH2 – pH1)/(E2-E1) is the electrode slope.
Thank you Mr. Cornel and Mr.Michel
First, find out the slop of the graph you are getting for E vs pH value.
if the relation is linear then proportion will help to find pH for any mV value available, but if it is not so then need to adopt some other methods.
Can any one help me to read TTL signal(5V TTL-Compatible) to read with Microcontroller ?I am having Pressure sensor with output TTL. Sensor having 3 pins One is Vcc ,GND and Signal(5v -TTL...
02 March 2019 2,665 5 View
07 August 2017 6,594 2 View
07 August 2017 7,071 2 View
07 August 2017 9,530 3 View
To measure DO in mg/L ,need to determine temperature ,salinity and pressure .I am using DO sensor which has temperature compensation also .Could you please help me to calculate DO by considering...
01 January 1970 4,657 1 View
How to calculate the RMSD values for a MD simulation using MOE?
07 August 2021 0 0 View
When I tried to energy minimization my system, I got fatal error as below. Fatal error: Atomtype opls_116 not found Although I've already added this line: ; include water #include "oplsaa.ff/spc.itp" to [molecultype] directive in my topology.
16 June 2021 0 0 View
effect of double bond on the flexibility of polymers
07 April 2021 0 0 View
I am writing to request some information for making polymer color flakes. I would appreciate the help.
17 March 2021 0 0 View
When I conducted the degassing experiment using decompression under anaerobic condition(and I used ferrous based chelating agent), the oxidized ferric were observerd. So, could you tell me the...
03 March 2021 4,892 5 View
I am using a Sutter electrode holder for my patch clamp experiments, but I'm having trouble keeping positive pressure inside my electrode due to a leak out of the back of the holder. I have added...
02 March 2021 5,829 1 View