I have a DLS result (done with repetition) of water dispersion of PEG coated nanoparticles using Micromeritics Nanoplus particle analyser where cumulant diameter was shown as 257.5 nm with PDI 0.545. However, if I check intensity and number distribution separately, I see that their Gaussian fitting gives peak at different positions. This is understandable as they are obtained based on different physical principles and detection methods. I also assume that my sample has formed agglomerates in water with different sizes. Now, while presenting in paper, which peak should I consider as the average hydrodynamic size? According to some instrument manuals, intensity peak should be considered especially when the sizes are not too small. Secondly, which PDI should I mention-the one obtained from the cumulant result (mentioned above) or the one obtained from Gaussian fitting of intensity distribution plot using the following relation?

PDI = ((Standard deviation)/(Average diameter))^2

I can get the standard deviation from the σ value of Gaussian fitting. This method gives a PDI value of 0.08.

Can I present average and PDI value from both the distributions using the above equation?

More Kashmiri Deka's questions See All
Similar questions and discussions