Hello,

I have run into an issue that has sparkled a debate at work: is the LLOQ/ULOQ impacted by injection volume under the following circumstances:

1) Same dilution factor of all samples

2) Volumetric deviations corrected for by internal standard (separate compound from analyte, not labelled)

3) Matrix effects studied and found to be negligible at all injection volumes

4) Good linearity of the regression model; not impacted negatively by inclusion of standards injected at lower volume

The problem arose when I, instead of performing a dilution, chose to rely on the correcting function of the internal standard and injected 1/5:th the volume (0.2 µL rather than 1 µL) of my most highly concentrated samples and standards and evaluated them as part of the same calibration curve as the samples injected at 1 µL. As I mentioned, matrix effects were investigated at all concentration levels as well as injection volumes and they were very low.

The software settings I use employ the "signal response" which is [area analyte/area IS]-ratio and the regression model from the calibration curve to calculate concentrations. Hence, under the circumstances I mentioned above the response is independent from injection volume - providing one does not drop below the background level or saturates the detector.

We ended up in a situation where one of my colleagues claim that the LLOQ for the samples injected at 1/5 should be adjusted and multiplied by the factor difference in injection volume (i.e. x5) and the ULOQ to be divided by the same factor. Another colleague claim it should be done the other way around.

In contrast to both I claim that the volumetric correction by the IS has already compensated for this and that LLOQ/ULOQ should remain at the set concentrations of their respective standards.

Can anyone help me out here?

Best regards and thanks in advance,

// Karl

More Karl Hansson's questions See All
Similar questions and discussions