Could you give more information on your experiment ? What detector did you use to follow the desorption of NH3 (TCD vs mass spectrometer), at what temperature do you observe this extra peak and what metal did you deposit ?
If you are using a TCD detector you cannot be sure that what desorbs at 753 K is ammonia, and hence that you indeed have an increase in the number of acid sites. You probably should try to perform the measurement using a mass spectrometer for detection. I have another question, during you NH3 TPD experiment, at what temperature did you pretreat your sample prior to expose it to ammonia ?