I used capillary tubes to do thermal death time test, and it is very difficult to get the sample (40uL) out of the tube. I tried pipette it but still small amount of liquid inside. Any ideas to get the sample out the tube in a more complete way?
Jie, I am aware of 2 options but not sure how convenient it is for your samples. 1. cut the sealed side of the tube to release the sample, you can retrieve it with a pipette. 2. place the tube in a falcon tube (or Eppendorf tube if your sample is small enough) and do a spin in a centrifuge it should take your sample into the larger tube and you can move from there DO THIS AT LOW RPM for safety.