The gamma constant for Co-60 is 0.152 mGy/h at 30 cm per mCi. 5 mCi will be 0.912 Gy/h. Divide each dose level by 0.912 for the number of hours exposure. The dose rate will change with distance by the square of the distance so dose rate will change.
Dose to thin films will be affected by material before or after the film. If the film is large in area, the dose rate will change over the film depending upon distance from the source.