I measured BET of few layered graphene.But, it is giving very low specific area around 8-10 m2/g only. Before BET measurement is there any process required for materials..?
Did you use an oxidation process to create your graphene? Run a TGA to check if you have fully reduced the material. If not, do thermal reduction in inert atmosphere or Ar/Hydrogen, then measure again.