I want to determine the latent heat enthalpy and melting point of natural Beeswax in order to develop an iso-thermal heat block system (620C - 640C). The TGA-DSC measurement for the off-shelf beeswax showed a linear increasing relationship on Q Vs T.
What could be the potential source for this linear increment deviating from the flat/baseline? Can the deviation be attributed from the instrument calibration fault or sample? such as long calibration of DSC - thermal drift - or sample's property such as thermal conductivity linearly increase with temperature or impurities?
#DSC #Phase_Change_Material, #TGA, #Theramal Conductivity #Enthalpy #Materials_sceince