I wanted to know if the satellite feature of W 4f appears within the 30 eV - 40 eV range such that it overlaps with the fundamental peaks. Some references, if you have them handy, will also be helpful.
Dear Indu Anjaly,
From Thermo database, the loss feature for W metals appears at 37eV, and 42eV for WO3. The W metal loss feature may overlagp WO3 4f peaks.
Regards,
Emile
I want to know more about Uranium ore deposits in world.
11 August 2024 6,720 0 View
I want to know more about diamond ore deposits in world.
11 August 2024 2,167 1 View
We assume that the difference is huge and that it is not possible to compare the two spaces. The R^4 mathematical space considers time as an external controller and the space itself is immobile in...
10 August 2024 6,678 14 View
If Banks do not provide credit facility, what are the options available for FPOs and impact on producer’s income?
10 August 2024 8,198 5 View
I used eye tracking to examine how participants from two different populations (A and B) react to an image. Participants in population A exhibit larger pupil sizes over time, but they also have...
10 August 2024 3,229 0 View
What are a “Farmers Producer Organization” (FPO) and its essential features?
10 August 2024 477 5 View
I have been doing the m6A dot blot for a while with no improvement, I am extracting the RNA, and I can see the dots although the three biological replicas give a different reading on the memberan...
10 August 2024 8,539 5 View
How do interactions between the biosphere, the carbon cycle, and the water cycle impact global warming and interaction between the atmosphere and the hydrosphere?
09 August 2024 3,291 2 View
I have input a moment load in module load Abaqus, i put my moment load on the node surface (using reference point). I have define moment in history output and make a set for moment too. But the...
08 August 2024 4,831 4 View
How is energy cycled through the Earth's climate system and how do matter cycle and energy flow through the rock cycle?
08 August 2024 8,162 0 View
Do these peaks in the curve indicate primer dimer or non-specific products?
01 August 2024 7,466 0 View
Hello. Thanks for your consideration to see my question. Recently, I conducted XPS anaylsis of g-CN that is prepared from thermal polycondensation of DCDA, so-called conventional bulk-g-CN,...
30 July 2024 9,824 2 View
Hi everyone, I'd like to better separate two close peaks coming from a bioconversion sample, I'm currently using a C18 KromaPhase 4,6mm Ø.I. SCHARLAU, particle size (µm): 10, pore size (Å): 100,...
30 July 2024 435 3 View
I'm using a InVia Renishaw Raman microscope system (785nm wavelength, 600l/mm grating density) to acquire Raman images on human brain samples, at 50um resolution. The samples are placed on MgF2...
30 July 2024 8,771 3 View
Dear all, I have noticed, as the ZnO content increases, the Raman intensity at 100cm-1 also increases. However, there is a peak nearby at 130 cm-1, which I could not find in any paper related to...
28 July 2024 5,410 1 View
These datasets will be used in the training of machine learning algorithms. Does anyone know any available data?"
17 July 2024 6,519 3 View
I do not know much about FTIR! The material Benzethonium Chloride USP was compared to a reference standard in regards to FTIR, as per the USP testing: Identification B. Infrared absorption (FTIR),...
16 July 2024 6,918 7 View
I have calculated the peak area from the XPS data. From this data how can I calculate the wt% of the material.
14 July 2024 6,951 1 View
Dear all, I'm currently working with CHI potentiostat and performing Impedance. However, for the circuit fitting the instrument must be connected to the computer. Since I prefer to analyse the...
10 July 2024 8,880 4 View
I have to compare the FTIR of the two samples, one is in powder form and other one is the powder dispersed in IPA or water. Both are same material, so how the bond will change
09 July 2024 8,011 2 View