It depends on the purpose. When it comes to the resolution between neutrons and gamma rays and the determination of particle energy, stilben is unsurpassed. When it comes to mechanical resistance, a plastic scintillator, eg EJ-299-34. If the criterion is detection volume, then a liquid scintillator, for example NE-213 (but it is sensitives to oxygen contact and is poisonous).
Note for NE-213 - it is cheap, but it needs to remove oxygen from it (eg by bubbling with argon) and has a considerable thermal expansion - it has to be encapsulated so as to create a space where the liquid can stretch out.
Note for stilbene - can not be produced more than about 50x50 mm and suffers from optical anisotropy, which is not a problem for detection (for spectroscopy yes).
Note for fast neutrons - these include neutrons with energy aproximatly ranging from 500 keV to 18 MeV.
أما بالنسبة للنيوترونات السريعة فإنه يفضل الكشف عنها باستخدام البروتونات المرتدة عند تشتت هذه النيوترونات على الهيدروجين . ولهذا الغرض تجهز البلورة في شكل خليط من حبيبات كبريتيد الخارصين ZnS والشمع لاحتوائه على نسبة عالية من الهيدروجين . وتعتبر هذه البلورة من أنسب البلورات للكشف عن النيوترونات السريعة .