We are running GC/MS analysis and have notices that our mid-range values in the calibration curve are showing as saturated when integrated. the height of the signal is in the order of 3E06 and the peak width is about 5 seconds. Our current sampling rate is 5 points per second at a scan speed of 1562 and cycle time of 190, with a step size of 0.1 m/z. Our mass range is 40-300. What are we doing wrong? (the only solution I have for it now is smoothing which solves the saturation problem, but drops the intensity of the peak and makes it wider)