1Wilford Hall Medical Center, Lackland Air Force Base, Texas, USA. [email protected]
Abstract
After observing a high incidence of low positive hepatitis C virus (HCV) antibody screens by the Ortho-Clinical Vitros ECi test (Orthoclinical Diagnostics, Raritan, NJ), we compared results against those obtained using another chemiluminescent analyzer, as well as two U.S. Food and Drug Administration (FDA)-approved confirmatory methodologies. To ascertain the true anti-HCV status of samples deemed low-positive by the Ortho-Clinical Vitros ECi test, we tested samples using the ADVIA Centaur HCV screen test (Siemens Medical Solutions Diagnostics), the Chiron recombinant immunoblot assay (RIBA) test (Chiron Corp., Emeryville, CA), and the Roche COBAS Amplicor HCV qualitative test (Roche Diagnostics, Indianapolis, IN) in a series of studies. Of 94 specimens positive by Vitros ECi, 19% were observed to be negative by Centaur. A separate study of 91 samples with signal-to-cutoff (s/co) values less than 8.0 showed that all but one was negative for HCV ribonucleic acid (RNA). In comparison with RIBA, 100% (77) samples positive by the Vitros ECi test with s/co values less than 12.0 were negative or indeterminate by RIBA. A final study comparing all four methods side-by-side showed 63% disagreement by Centaur for Vitros ECi low-positive samples, 75% disagreement by RIBA, and 97% disagreement by polymerase chain reaction (PCR). In conclusion, the Ortho-Clinical Vitros ECi Anti-HCV test yields a high rate of false-positive results in the low s/co range in our patient population