I am trying to determine OIT of PE using DSC and TGA. Process is heat 5 mg, 1mm thick unstabilized HDPE on open Al Pan under Nitrogen (200 mL/miniute, 50 psi, introduced in DSC) to 190 C, maintain for 5 minutes and switch to Oxygen (50 mL/minute, 50 psi pressure). I am not seeing shift in base line or increase in exotherm . Iw as expecting OIT of 5-10 minutes for unstabilized HDPE but I do not see it even after 180 minutes. Something wrong with test. However, rotometer or both gases are stable indicating no gas leaks.
Is 50 psi too low a pressure? is 50 mL/minute not enough?