Hello All researchers of Health/Medical/Radiation Physics, at what Dose Rate of Co-60 teletherapy source should be replaced? I have searched but find NAND!
Is any IAEA, AAPM guidelines/recommendations available?
The note: for Co-60 tele therapy sources in external radiotherapy , the maximum dose rate don't should lower than 70 cGy/min ( in SSD=100 cm & 10 x 10 cm. x cm. field size ).
excuse me. The note: for Co-60 tele therapy sources in external radiotherapy , the minimum dose rate don't should lower than 70 cGy/min ( in SSD=100 cm & 10 x 10 cm. x cm. field size ).
Yes, minimum dose rate 50 cGy/min at SSD=80 cm for Field size= 10 X 10 cm x cm is acceptable for external beam radiotherapy. Because below this dose rate the treatment reproducibility will decreased and treatment outcomes will be affected.