Take Glass-fiber filter (Whatman 934-AH , Cat No 1827-047).Wash it @ three successive 10 ml vol of reagent grade water. Take sample volume 10-200 ml depend on turbidity . pass sample with vacuum pump. remove the filter paper. Dry@ 103-105oC in oven,cool in desiccator weight it.
mg total Suspended Solids /L=(A-B)x1000/ sample volume,ml
Total suspended solids (TSS) can be accurately measured by filtration through 0.45 Mili micron membrane fibre.. Detail about analysis is well describable in published literatures.