I am using GC 2010 of Shimadzu. In my analysis I am using TCD Methanizer FID. Problem is, what could be the possible problems if I keep the Methanizer at 380 C the whole night, with the total flow at 10 ml/min?
I got the information from one researcher who used the methanizer before. There is no problem if I keep methanizer onat 380 C whole night even for whole week. But I have to keep flow in the system. It is highly recommended not to change injector septum without keeping down the methanizer temperature at ambient. (generally take 2 hr).