What are the factors for misfire for Bomb Calorimeter to happen? Recently I've noticed that this equipment can't give a result because of misfire during the operation. And I've noticed that the electrode (both electrodes for holding the fuse) are a little bit oxidized. After I clean it with sand paper and sonicate for 30 minutes then the operation going very smooth.

So my question is, does electricity flow on the surface of the electrodes or in the interior? Will the Purified Oxygen can cause the electrode to be oxidized very fast?

Model: Parr 6400