Hello and good day.

I am using a Q Exactive Plus and in the experiment pane for a Full MS / ddMS2 (TopN) there are three settings for the automatic gain control (AGC), the number of ions to be in the detector Orbitrap. They are labeled as such in the Full MS submenu, dd-MS2 / dd-SIM and dd Settings (minimum AGC target) as marked in the attachment.

The first one is for Full MS, the second I guess is the threshold that indicates if its selected for MS2 when there is not an inclusion list and the third one when there is an inclusion list, is this correct?

My goal in this case is to use an inclusion list. In my last run some targets were not acquired for MS2 despite being clearly present in MS1. Decreasing dd AGC minimum (the third setting) helped for some (1e3 down from 8e3) but not all, and in the TIC trace the missing ones are above 1e3. What can I do to force these missing ions into MS2?

Thank you!