Voltage is 20kV with 95mA.
Prob size large (18).
Jeol 7000f. Aztec software.
Dear Reza Taherzadeh Mousavian ,
Some ideas:
There may be an error on the detector side.
a) Either there may be no voltage applied to the detector, or
b) the detector voltage is too high...
c) low level discriminator (LLD) set too low?
With respect to the acquisition parameters you gave above:
lower voltage and current significantly and see what's happening.
Maybe sparking is happening somewhere...(is the sample sufficiently grounded?)
Good luck and
best regards
G.M.
Thanks Gerhard. I think current might be too high and I'll check it. Do you know what is OL Aperature?
to my opinion, OL stands for objective lense.
The problem was easily solved. The camera was ON! :)
@GERHARD MARTENS
Thanks Reza Taherzadeh Mousavian ,
for feed back.
Fortunately some problems/challenges will solve themselves...
I have the same problem on Axia Thermo Fisher. What means "The camera was ON"?
Aleksey Yakushkin Thanks for your message. When your camera for observation of the sample holder inside the chamber is ON then you won't get true EDS results. You should turn it off to have dead time around 45-75 percent.
Here's a serious question: If we were to stop emitting all greenhouse gases right now, would the Earth's temperature start to cool down, or would the existing greenhouse gases continue to warm the...
30 July 2024 5,123 4 View
final solution volume is 100ml
03 June 2024 417 0 View
could you please help with a protocol? should i do it directly on the aluminium stub after dehydration with 100% alcohol or not? should i release the pollen or whole anther is better and dehisce...
03 June 2024 4,793 8 View
SPK tool steels are used for cutting wires made from low carbon steels. We are searching for a cheap and effective method for increasing the cutting characteristics of the cutters.
14 May 2024 4,948 0 View
Hello I recently encountered a problem with the reciprocal of step size plot in Comsol Multiphysics. After a certain time (seconds) in a time-dependent solver, the plot oscillations will be...
04 May 2024 7,278 0 View
Hi everyone I produced a batch of virus , enriched it with centrifugation, and titrated it with T cells after 48h. For my experiments I want to use MOI 3 and that's equal to 1/12 dilution. but at...
02 May 2024 9,727 3 View
InfoScience Trends offers comprehensive scientific content that delves into various facets of information science research. This includes but is not limited to topics such as information...
26 April 2024 9,521 1 View
I work in media literacy field, Cognitive Science field, Cyber Governance field and Cyber policymaking field. I asked introduce me experts in fields that I mentioned.
06 April 2024 6,255 0 View
Garret, 2020 used traps of blue, yellow, and white colors for sampling and collected several specimens.
06 April 2024 6,995 2 View
What are the laboratory procedures for identifying the concentrations of formic acid and acetic acid in aqueous solutions using titration, spectrophotometry, or gas chromatography methods...
06 April 2024 187 3 View
CK, OK, NAK MEANING IN SEM EDAX GRAPH
02 October 2023 8,337 6 View
Dear Experts, I would like to understand the analytical value interpretation of my sample which is subjected for elemental characterization study by EDAX SEM analysis. The reported results...
02 June 2023 9,129 2 View
What are the possible reason behind the presence of unwanted elements which is predicted during EDAX analysis
17 May 2023 2,389 7 View
I deposited around 50 nm thick FeGe film on a silicon wafer using a DC magnetron sputtering. The compositional analysis data from FESEM EDAX at the deposition area shows that the oxygen atomic...
17 November 2022 4,036 8 View
EDAX ZAF Quantification (Standardless) Element Normalized SEC Table : Default Elem Wt % Chem K-Ratio Z A C K 52.50 5.92 0.1691 1.0276 0.3133 1.0002
31 October 2022 9,582 2 View
We have done the Energy Dispersive X-ray Analysis (EDAX) of BiOCl, but other than this method which method is the best to determine the elemental composition of BiOCl?
11 September 2022 9,930 1 View
I have rice husk and ash samples and want to evaluate them quantitatively for elemental composition. I would like to compare the results between samples and choose the best sample. So I would like...
11 January 2022 7,619 5 View
i need to get characterization reports for my synthesized manganese oxide nano particles.
21 May 2021 9,996 5 View
I have come through a few literature,were the presence of particular clay mineral has been validated through the Si:Al ratio obtained from EDAX analysis. But, I would like to understand the basic...
08 January 2021 6,299 6 View
Regarding EDAX data analysis Is it possible to change these wt % values in to numerals to do the statistical analysis. If so, how to do? Can anyone please tell? Calcium: Wt% (52.41) and At %...
24 August 2020 6,170 3 View