I want to generate second harmonic using the long-wavelength (1600nm) part of the supercontinuum as pump. How to decide on choosing the right focal length lens for SHG. I am using 10mm-long MgO-PPLN. The average power at the output of supercontinuum is ~100mw and ~40mw for SHG (after long-pass dichroic mirror). Since the temporal profile of the signal after SC is not known (specifically for the 1600nm part ), is it suitable to treat that signal as continuous-wave for SHG pump. For CW pump, the literature suggests that the ratio of crystal length to the confocal parameter (2*Rayleigh range) should be 2.84 for optimal SHG. From where we may calculate the right focal length of the lens to focus light into the SHG crystal.