I'm currently experiencing some challenges while trying to determine the atomic percentage of elements from XPS survey spectra. I'm using OriginPro software for my analysis.
Please be always sure to cite which set of sensitivity factors you are using, there are also different ones. I'm not saying you need to switch, but for reproducibility the set has to be specified.
When I look at the baselines, it seems to me they are cutting a little into the flesh of the peak on the left side each time, so please make sure that doesn't happen. I am not in fundamental opposition towards using Origin for XPS evaluation, but you have to be a lot more careful it does the right thing than when you're using a dedicated XPS software,
Edit: at second glance, have you tried putting one big Tougaard line through the while spectrum? I think the normal way would be to subtract the peaks individually.