What are the factors affecting the measurement of electrolytic conductivity using DC and AC techniques? What are the challenges and precautions related to electrolytic conductivity measurement? How to compute the cell constant?
Electrolytic conductivity measurement is a key technique used in various fields to determine the concentration of ions in an electrolytic solution. It applies a voltage to two electrodes immersed in a solution and measures the resulting current. The ability of the solution to conduct an electric current is related to its ion concentration.
DC (Direct Current) Technique
A direct current is applied across the electrodes in a DC measurement, and the resulting potential difference or current is measured. One of the main issues with DC measurement is electrode polarization, where charges build up on the electrode surfaces, resulting in a concentration gradient that can skew the measurements.
AC (Alternating Current) Technique
The AC method addresses the problem of electrode polarization by using an alternating current instead of a direct current. Because the direction of the current is continually changing, charge buildup is minimized, and more accurate conductivity measurements can be obtained.
Factors Affecting the Measurement
Temperature: Conductivity is highly temperature-dependent, so it is crucial to control the temperature during the measurement or correct the results for temperature effects.
Electrode Spacing: The distance between the electrodes affects the measurement, and this effect is accounted for using the cell constant.
Concentration of the Electrolyte: The concentration of ions in the solution directly impacts the conductivity.
Ionic Mobility: Different ions have different mobilities, which impacts their ability to conduct electricity.
Electrode Material and Cleanliness: The electrode material can impact measurements, and any contamination on the electrode surfaces can also skew results.
Challenges and Precautions
Polarization Effects: As mentioned, electrode polarization can significantly impact measurements, especially with DC techniques. Using AC techniques or specially designed electrodes can minimize these effects.
Contamination: Contamination of either the solution or the electrodes can significantly skew measurements. Cleaning and maintaining the electrodes and using high-purity solutions where necessary are crucial.
Temperature Control: As conductivity is temperature-dependent, temperature control during measurement is critical, or a temperature correction factor must be applied.
Computing the Cell Constant
The cell constant (K) is a factor used to account for the geometry of the electrode cell. It is typically determined by calibrating the cell with a solution of known conductivity. It is given by:
K = d / R
where d is the distance between the electrodes, and R is the resistance measured using a solution of known conductivity. The units of the cell constant are cm-1.
It's crucial to note that if the electrodes are not parallel plates (often in practical applications), the cell constant must be determined through calibration rather than simple geometric calculation. It's typically a good idea to periodically recalibrate the cell constant to ensure accurate measurements, as it can change over time due to factors such as electrode wear or buildup.
I received got better response to my query, I think contamination from environment gases with electrolytes may also be one of the factor Rana Hamza Shakil