23 October 2018 18 2K Report

I'm interested in quantifying the amount of short-chained (C4-C6) aliphatic esters. I work on an Agilent GC-MS system equipped with a Gerstel thermal desorption unit and a liquid nitrogen cold injection system.

The "standard" setup I use includes heating of the TDU to 300°c while keeping the liner (glass wool only) at -100°c. I normally use a non-polar column (db5).

The highly volatile compounds are detectable, but the peaks are sometimes very deformed. Also, the highly volatile compounds sometimes come out in two adjacent peaks.

Does anyone have tips on how to optimize the system to quantify these compounds? Polar column? Different liner? Colder injection system?

Thanks!

More Omer Nevo's questions See All
Similar questions and discussions