I want to the reason for getting only one band
Manai,
this should be due to degradation. The thing is that this degradation may occur also during electrophoresis.
Goldmann,,
this mean that the RNA may be ok
So what i can do to prevent degradation during electrophoresis.
Thanks in advance,,
MAnal
Dear Manal,
Did you compare your band with ladder? In which range does the band assemble? Maybe all the RNA is degraded and the band is caused by some contaminations.