I am working on e-cigarettes and e-liquids as a part of my masters project.
Its boiling point is 247 deg.C. It will be hard to move it to headspace phase at high rate.
In what temperature are you thermostatting your sample (and how long)?
You can try to increase the temp or use additionall techniques like SPME od SDME coupled with headspace.
regards,
GB
Grzegorz Boczkaj I am doing it at 250 deg. C .
But i anyway managed to find it with just GC-MS.
Thank you.
Regards,
Meg
Nicotine has good MS response. But why do you do headspace analysis? I suggest direct injection (e.g. solvent extracts after LLE/DLLME etc.) of Your sample using split/splitless injection mode (of course it depends on Your matrix).
Best regards - Mateusz
Since machine language and AI are the technologies of the future I want to know hoe we can use both of these for improving project management.
11 June 2023 1,368 15 View
Is there any theoretical way of finding it?
21 June 2020 6,735 0 View
For example: Wet soil sample weight was 42.5 g same sample oven dried at 105 C weight 31.0 g, so the I come to know there is 37% water content. How much additional water I should add to increase...
26 May 2018 3,879 9 View
Cell has as area of 3*3 cm2 .0.2mg/cm2 is the loading of Pt on C
31 March 2016 3,598 4 View
The catalyst after loading on GC needs to be studied for OER,ORR.
08 November 2015 9,490 9 View
Project Management Institute notes that 81% of professionals say that AI is impacting their organizations. That number is likely to increase further in the coming years. Automation holds immense...
01 January 1970 100 1 View
Could you recommend some articles on Urban Transportation System optimization and Innovation?
13 August 2024 2,595 3 View
A crude extract of fungal culture using EtOH was subjected to column and TLC and partially purified compound was obtained. UV vis spectrum of the compound/s has max absorbance at 218nm. The...
11 August 2024 9,801 2 View
Can anyone explain this method? Especially the last statement where it says only at 1.5 to 2.5mins was the MS/MS connected to the UPLC. How is that possible, is it a feature in this specific...
11 August 2024 8,141 3 View
Hello experts, Does anyone know any free software about retention index prediction ?
08 August 2024 7,403 2 View
I'm currently working on calculating the collision cross section (CCS) for various ions, and I'm facing challenges when dealing with sodiated and multiply charged ions. Most of the resources I’ve...
08 August 2024 8,329 0 View
I'm working on selecting antibodies against a recombinant protein that has a His-tag. My idea is to first bind the recombinant protein to a HisTRAP column and then use this column for an affinity...
07 August 2024 505 3 View
Dear QE-users, In the method where full MS positive mode and PRM mode are used, we always get an incorrect auxiliary gas reading (41 instead of 25). This only happens in this method; other...
06 August 2024 4,953 0 View
Hello, Can anyone provide me with the absorption coefficient of methane gas at 7.7 um? Any reference?
06 August 2024 980 5 View
There are anaerobic gas pak jars and environment providers in the laboratory that provide anaerobic environmental conditions for the growth of anaerobic bacteria. However, there is no anaerobic...
05 August 2024 4,945 3 View
Why is TLC not a form of partition chromatography and paper chromatography not a form of absorption chromatography?
03 August 2024 7,035 3 View