I am running a test to detect antibodies to myeloperoxidase (MPO) in patient sera. When I use different lots of wash buffer on the same test kit, the reactivity changes significantly, sometimes over 50% less reactive.
The test is run on a magnetic bead with MPO attached to it. Patient serum is applied to the beads, decanted, and then washed 3 times with wash buffer. Isoluminol is used to detect antibodies that are left on the beads after the washes. I have already ruled out all of the other components in the test except the lot of wash buffer that is used. Generally, it seems like older lots of wash buffer lead to less reactivity, but the trend is not very strong.
The wash buffer consists of 10mM potassium phosphate buffer, 120mM potassium chloride, 0.05% Tween 20 (detergent), and 0.09% sodium azide.