04 March 2025 3 9K Report

Hey everyone,

im struggling to explain whats the cause of the following phenomena:

We are using HPLC to validate the purity of our compounds. So in the best case we obtain chromatograms with a single peak an maybe some minor impurities. However, i started noticing, that only in the case of fluorescein compounds multiple peaks are appearing in the chromatogram. We are using a DAD detector and the spectra of these "duplica" is all the same, indicating is should be the same compound. Also NMR of the sample shows a single, clean product. All other measured substances are showing just a single peak. Does anyone has an explanation for this behavior?

The column is a reversed phase C18 column

Im running a gradient from 0 to 100% B

A= 5% ACN and 0.1 % MSA in Water

B= 5% Water and 0.1% MSA in ACN

Detection in the chromatorgam shown is 210 nm

thanks,

Jan

More Jan Brahm's questions See All
Similar questions and discussions