I'm using ammonia as a reagent gas for chemical ionization MS. Set at 2 bar, the pressure on the pressure reducer drops during a series of analyses to -1 bar, resulting in a drop from 1.5 mL/min flow to a steady 0.4 mL/min. Of course, this non-optimal flow yields serious losses in sensitivity and inconsistent measurements.

The pressure reducer is suited for ammonia and two types have been tested, the tubing has a large amount of windings stacked in a vertical position and was replaced by a longer one, and even the entire module with mass flow controller and valves inside the MS was replaced. No results. The technicians from both the MS company and from the gas supplier do not see the cause.

Has anyone experienced similar issues, or does anyone have an idea about potential solutions to this problem?

Similar questions and discussions