Dear Atefe,
When the problem arising, during or prior to analyses?
There may be several reasons for that:
1. The condition of the quality-control beads (flow check etc.): usually inappropriate mixing;
2. Gate issue - adjust the gate
3. inappropriate flow rate
4. Pressure issue - try to clean the flowing tubes, remove bubbles;
5. The laser problem:
a) fluctuations of temperature (stabilize the temp., restart the instrument)
b) fine manual adjustment of the laser may be useful (for example you can do it on FC500 Beckman).
Thank you
Hi, Eleni.
Yes, generally, the issues like CV deviation you need to check prior to data collection, cause they may affect your further analyses.
Some issues, however, may exist during analysis (sample preparation).
This is why I`m asked if the problem happens during or prior to analyses.
Dear Atefe
I guess you shoud be careful about chopping the samples. Also it is necessary to exam your samples in different gain (channel) of FCM instrument.
Best wishes
If QC of the instrument shows CV with in ranges, then there is something with your sample. First acquisition of sample has to be done on low or medium pressure. as you increase the differential pressure, it increase the CV.
Thank you for your good point.
09 October 2016 8,724 2 View
06 July 2016 5,742 8 View
10 November 2014 1,631 9 View
See above
08 September 2014 7,016 8 View
I want to use from this software for analysis of obtained peaks from flow cytometry.
08 September 2014 556 3 View
Hiiiii everyone! I have an enquiry on statistical analysis. I was looking for many forum and it's still cannot solve my problem. I want to compare means of two groups of data but only with two...
03 March 2021 8,796 3 View
I am on the lookout for the Enhanced Yellow Fluorescent Protein (Aequorea victoria) DNA sequence. Does anyone know where I can find it? Thank you in advance
03 March 2021 3,568 1 View
Hi, I want to start testing pitfall trap to obtain ants samples, but I need to conduct molecular analysis on those insects. So, what kind of fluid can I use? Ethanol expires too early and I need...
03 March 2021 5,978 5 View
What's the best way to measure growth rates in House sparrow chicks from day 2 to day 10? Since, the growth curve from day 2 to 10 won't be like the "Logistic curve" it might not follow logistic...
03 March 2021 1,401 3 View
I have conducted and published a systematic review and meta-analysis research with the topic related to public health and health pomotion (protocol was registed in PROSPERO). Now we want to...
03 March 2021 8,920 3 View
dear community, my model is based feature extraction from non stationary signals using discrete Wavelet Transform and then using statistical features then machine learning classifiers in order to...
03 March 2021 6,994 5 View
I just wanted to check if I need to run a linear regression separately if I am using PROCESS MACRO to run mediation analysis. Thank you.
02 March 2021 4,359 3 View
If the detection range is in ng/ml but the reference range is in ug/ml for a molecule or protein in serum or plasma .how to dilute and what is the initial volume to be taken for quantitative analysis
02 March 2021 7,670 3 View
Is There Any Feasible Method To Test The Efficiency Of Fluorescent Compounds Other Than UV Spectrometers ? Suggestions Would Be Appreciated !
02 March 2021 5,785 3 View
Dear all, I just saw this example in FB on the difference between laminar flow and turbulent flow. The flow from a pipe hits the sink and scattered. The flow from the pipe is characterized as...
02 March 2021 1,999 5 View