Dear Lee Jieun you have negative points and data also not reliable .Please do the baseline correction and carryout the experiment again. Transmittance can not be more than 100 % .
Normalizing can solve the problem of the negative transmittance issue, but the original problem is not solved yet.
By the way, I found the problem by monitoring energy and interferogram yesterday. Energy value is zero, and interferogram spectrum is like the picture I upload. The normal case is that the energy value has the number and interferogram main peak has to be set to the middle. Does anybody have another idea about this problem?