Dear Sir. Concerning your issue about the protocol/SOP for validation of a steroid harmone done in LC-MS. High performance liquid chromatography-tandem mass spectrometry (HPLC-MS/MS) has the potential to find its place in the clinical laboratory medicine for quantification of steroid hormones. A prerequisite for the application of a new analytical procedure in clinical diagnostics is standardization to minimize analytical intra- and interlaboratory variability and inaccuracy. The sample preparation is performed by solid phase extraction (SPE) on a 96-well plate. The standardized assay has been validated for human serum in terms of lower and upper limit of quantification (LLOQ 0.01-32 ng/mL, ULOQ 5-8000 ng/mL), linear correlation coefficient of calibration (R(2)>0.9966), intra- and inter-day precision (intra-day 1.1-8.8%, inter-day 5.2-14.8% and 8.2-18.6% for 11-deoxycorticosterone), accuracy (intra-day 88.3-115.5% and 109.3-128.2% for 11-deoxycorticosterone, inter-day 91.4-117.2% and 102.3-137.1% for 11-deoxycorticosterone), analytical total error (3.6-17.8%), proficiency test accuracy (85.4-113.4%), recovery (68-99%), and metabolite stability (freeze/thaw stability 95.5-108.1%, short term stability 86.9-107.2%) . Steroids were extracted from serum or plasma using automated solid-phase extraction where samples were loaded onto C18 columns, washed with water and hexane, and then eluted with ethyl acetate. Quantitation by LC-MS/MS was done in positive ion, multiple reaction-monitoring (MRM) mode with an atmospheric pressure chemical ionization (APCI) source and heated nebulizer (500°C). Deuterated steroids served as internal standards and run time was 15 minutes. Extraction recoveries were 87–101% for the 8 analytes, and all intra- and inter-run CVs were ≤8.25%. This quantitation method yields good recoveries with variable lipid-content samples, avoids antibody cross-reactivity issues, and delivers results for multiple steroids. I think the following below links and the attached files may help you in your analysis: