Hi Sumit, you can use either forms. For measuring powders you usually need front-face detection and maybe a double detection monochromator since you will have a lot of stray-ligth coming from the scattering produced by powder particles.
Measurements of bulk glass samples are maybe simpler, as you have less scattering; moreover if the sample is thick enough you can use right-angle detection which is the most common instrumental configuration.