I am using a HPLC method using a Mobile phase Water and Acetonitrile each consisting 0.1% formic acid. The lyamda Max was 256. It was working properly. However, now it shows unstable base line.
Thank you in advance
Paras mani Giri , many thanks for your message. What kind of unstable baseline do you observe?
pulsations? (Comment: This might be an air bubble in the pump.)
high frequent (white) noise? (Comment: This might be an air bubble in the UV-cell)
spikes (sharp only one second or less high signals) on the baseline? (Comment: this could be due to an older UV-lamp)
ad 1) purge the pump.
ad 2) flush the UV-cell with iso-propanol (e.g., using a syringe.)
ad 3) replace the lamp.
Any additional information you can give us is a bonus.
Long time ago, I was told that troubleshooting starts with information. The more we know, the more likely is it that we can find the culprit.
The questions are "W"-questions: What, when, who, why, hoW...
Good luck, and let us know how the troubleshooting went
Detlef.
Also did you ultrafilter the mobile phase? Is the mobile phase freshly prepared or sitting around a long time?
I am having an unknown bacterial hit so far, which is inhibiting the growth of multidrug-resistant bacteria. I want to characterize that antimicrobial compound.what should I do to characterize...
21 September 2020 6,869 3 View
Fake researches and predatory journals
14 September 2020 4,530 7 View
For SEM-based analysis of trichomes (soybean plant) can we preserve the samples collected for some period of time? or after following standard fixation procedures it can be preserved under low...
01 June 2020 5,665 1 View
Hi all : Please, I need some help on Heat Radiation Transfer: How I can derive the formula of the View Factor for a particular configuration, in a way of ending with an expression for the VF as...
15 May 2020 4,929 6 View
Hi all , How can I consider the Optical properties of a surface into the Stefan-Boltzmann law for a problem of Radiation Heat Transfer between a hot Body and a real Surface (namely, a surface...
15 May 2020 9,449 8 View
Dear all : I need to solve the following integral (attached as an image file) The context is on the calculation of View Factors in Radiation Heat Transfer I worked out the this expression from...
15 May 2020 7,500 16 View
Hi all, I'm trying to perform a calculation on the Flux due only to radiation, received at the surface of a Flat Plate Solar Collector placed on the surface of the Earth. The collector is tilted...
14 May 2020 604 24 View
Dear all : I'm trying to solve the following problem using COMSOL Multyphysics. I have an enclosed geometry (please see the attached images) with a top flat surface (the celing) and an undulated...
19 March 2020 7,455 5 View
Dear all: I need help, I hope there is someone out there who can help me. I finish drawing the 2D geometry of an Air Solar Collector with an undulated absorber plate (in contrast with a flat...
19 March 2020 5,935 9 View
Hello to all : This question is originated from the idea that every material (except for Superconductors) has a Seebeck Coefficient (S) different than zero, and from the idea that in every case...
19 March 2020 4,192 5 View
I am try to make the Paal Knorr reaction between 2.5 hexnedione and 6-amino hexanoic acid, my problem is type of solvent to use, because 6-hexaminohexanoic acid is soluble in water but I am not...
02 March 2021 4,443 2 View
Also when RHAMM binds hyaluronic acid, they get internalized, will RHAMM also be degraded? Or both CD44 and RHAMM will be transferred back to the cell membrane? Asking for breast cancer cell line...
01 March 2021 8,169 2 View
based on what I've studied on multiple papers, for peptide and protein identification Alpha-cyano-4-hydroxycinnamic acid (CHCA) is an appropriate matrix to use, I also know that the CHCA powder...
26 February 2021 6,799 3 View
I have set up a qPCR run on MxPro with SYBR green and ROX as a reference, however, when I go to the analysis section the program seems to only collect data for the dissociation curves and not the...
25 February 2021 8,039 1 View
I am currently trying to stain for PI3KBeta in MDA-MB-231 cells. My current issue is that 1) beta stains all over the cell and 2) there is an increase in signal at the edge of the cell and this...
24 February 2021 8,582 1 View
Hi! I would want to know if I can use acetic acid with an ELITE WAX Perkin Elmer column for Gas Chromatography. This column has a polyethylene glycol phase, but I can't find the manual to check...
22 February 2021 730 2 View
I have a fixed sample and would like to use the murexide reaction to test for uric acid composition. I no longer have fresh samples
22 February 2021 1,328 1 View
I notice that upon washing and air-drying of my RGO/SPCE surface and characterizing again via DPV, the current response towards ferricyanide probe significantly decreases. My worry is that the...
22 February 2021 5,324 3 View
I have used Picral etchant (4.2 g Picric acid, 10 mL Acetic acid, 70 mL Ethanol, 10 mL distilled water) but it didn't work. Is there any alternative?
21 February 2021 2,558 5 View
Hello! I am interested in determining whether a poly-unsaturated fatty acid I have engineered E. coli to produce is being incorporated into the membrane of the cells or if it is simply a free...
21 February 2021 7,447 2 View