for the calculation of magnetic susceptibility ,i need the value of mass magnetic susceptibility (χg) ,that will come if balance calibration constant is known .how to do that..give suggestion plz
Dear Ankit,
Calculation of Magnetic Susceptibility
in this table X = Chi
The volume susceptibility (Xv) = I / H Where:
I = Intensity of magnetism produced in a substance
H = Intensity of magnetic field applied
The mass susceptibility (Xg) = Xv / d Where:
d = density of substance
The calculation of Xg from the readings on the Magnetic Susceptibility Balance is simple
Xg = Cl(R-Ro) / 109m Where:
C = calibration constant of the balance
I = length of sample in cm (1>1.5cm.)
m = mass of sample in gm
R = balance reading for sample in tube
Ro = balance reading for empty tube
For more on this topic, please see the publication contained in the following link:
http://www.sherwood-scientific.com/msb/msbindex.html
Hoping this will be helpful,
Rafik
thank u sir , but i want to ask u how to know about that calibration constant of balance (C)and at the place of (C) what value i have to written
The calibration constant of balance (C) is usually 1 if it has been calibrated.
For more details, please see the following link:
https://en.wikipedia.org/wiki/Evans_balance
I have to learn the magnettic data fitting of variable temperature graphs and how to make input file to run this program? tutorial video give some more clear idea insted of literature to learn...
10 November 2018 3,203 0 View
CHN analysis , Mass , UV and CV .
07 August 2018 5,931 0 View
column chromatography is not possible here because of chance of degradation through hydrolysis and washing also not possible because of same solubility.how can i isolate them and to get my desired...
06 July 2017 6,834 9 View
Please find the file attached.
01 February 2017 7,964 5 View
i just gone through this article , see the attachment
31 December 2016 5,471 0 View
at the time of submitting samples for NMR or HRMS, we should confirm the solubility of compounds. Sometimes a compound is soluble in just adding the solvent and sometimes using sonication, it...
07 August 2016 4,323 3 View
i have prepared one manganese complex from N,N-bis(picolinyldene) hydrazine and 2-pyridinealdimine. in the absence of Mn(II) they do not react to yield the ligand. for more information i attached...
07 August 2016 5,825 4 View
i have tried to prepare metal free macro cyclic schiff base ligand number of times but i could not get it ,i m getting only polymeric type compound that is not dissolving in any solvent ,give me...
07 August 2016 9,007 4 View
Dear Sir/Madam, Does anyone know about the synthetic procedure for 2,6-Diformyl-4-methylthiophenol ? I have gone through a no of literature to find the synthetic procedure to prepare this compound...
07 August 2016 449 6 View
in case of organic ligand there r several techniques for purification,but in case of metal complex(Cr,Mn,Fe,Co) with schiff base i have not seen any literature for the purification of them except...
06 July 2016 1,419 8 View
Can anyone explain this method? Especially the last statement where it says only at 1.5 to 2.5mins was the MS/MS connected to the UPLC. How is that possible, is it a feature in this specific...
11 August 2024 8,141 3 View
Hello experts, Does anyone know any free software about retention index prediction ?
08 August 2024 7,403 2 View
I'm currently working on calculating the collision cross section (CCS) for various ions, and I'm facing challenges when dealing with sodiated and multiply charged ions. Most of the resources I’ve...
08 August 2024 8,329 0 View
Hi, I'm currently working on a project where I need to plot the atom-projected band structure using GPAW. I've been able to calculate the band structure for my material, but I'm having trouble...
07 August 2024 269 3 View
If you think of electrons with spin as bar magnets, you know bar magnets of opposite polarity as long as they're not occupying the same spatial location don't cancel out each other's magnetic...
03 August 2024 3,927 2 View
This question tackles the complex issue of ensuring that while AI can tailor educational experiences to individual needs, it doesn't neglect the equally important aspect of fostering social...
02 August 2024 4,751 3 View
I am using a Bruker 600M solid-state NMR spectrometer with a Micro 2.5 microimaging system. The test sample is a tube of 1M LiCl aqueous solution, and the nucleus detected is 1H. I am trying to...
01 August 2024 9,227 1 View
Following curve related to Co0.6Mn0.4Fe2O4 Magnetic nanoparticles with extra phase a-Fe2O3. I will appreciate to help me to interpretation this spectroscopy, especially the blue line.
31 July 2024 9,140 2 View
I want to perform a CRISPRa/i library screening for candidate regulatory factors that affect pathogenic Th17 differentiation in PBMC. But I am wondering how to define the size of gene library, so...
31 July 2024 2,150 0 View
EDX analisys
31 July 2024 5,947 2 View