I would like to resolve peak tailing in the hplc analysis of one drug. Mobile phase used is 100% Methanol. The sample injected after saponification. The sample pKa value is 16.5 before saponification. C18 column ,250 mmx4.6 mm, 5 um was used.
You are using a methanol mobile phase, on C18 column. For basic or acidic components you get a better result, with less tailing using an appropiate buffer.
Try a water / buffer mobile phase at slightly basic pH, depending on the pKa of your sample component of interest.