So, given the following, are my calculations correct?
- the dial for series resistance on the amp reads 0.3 (= 3 megaohms)
- the compensation dial reads 75%
- the maximal current at one point in time is 6 nA
- the command voltage is to -10 mV
Ru = 3 * 25 / 100 = 0.75 megaohms
Vm = (-10 mV) - (6 nA * 0.75 megaohms)
(after converting everything into volts, amps, and ohms)
Vm = -14.5 mV
And to your point of clarification - I can say then that the maximum voltage error for this experiment is 4.5 mV (since the 6 nA is the maximal current), but not that the voltage error for the entire sweep is 4.5 mV.