No, vice versum. Excessive illumination in not possible in UVVis spectroscopy. I think you are just getting 3A+ absorbance. Try dilution, recheck wavelength an make additional precautions if working in 180nm-210nm area
i don't think that the error is due to excessive illumination because the instrument works for the first measure without any problems and after that it don't work and appear the error "overflow".
Someone said me that i need to rename the method with a short name, but i think that this is a not scientific solution! I worked almost in all condition (opening the slits or closing them in order to avoid signal saturation) but with no results. I don't think that is a problem of my samples but i think it's an instrument problem.
I worked in 200-500 nm area with an excitation wavelenght =350nm and emission wavelenght = 450nm. The slits were both set at 4.0
Dear Irene, this may be caused by a hardware overflow. Can you set the number of points, resolution or whatever increases the size of the data or the aquisition rate? This may be related to an overflow in the aquisition cards or analog to digital conversion hardware. After the first aquisition of a data set to big to processed a second try results in the overflow. Another suggestion would be an I/O error.
There's nothing wrong with your equipment or your samples... :)
PerkinElmer fluorometers have a fluorescence scale from 0 to 1000 arbitrary units. Every time your sample has a fluorescence above 1000, you will get an overflow message.
The solution is simple: you can (a) use a lower slit value (2.5 nm, for instance), (b) use a lower voltage of the photomultiplier, (c) both. If your sample is very, very fluorescence you can always dilute it.
If those options didn't work, you can select an interference filter. Keep in mind that you have to select one with a wavelenght between your excitation and emission wavelengths.
The last option is to select the filter called "1% transmitance". This is the very last choice because it makes all your readings very low and you will loose sensitivity. I don't recommend to use this one.
Very important note: since fluorescence increases exponentially with the voltage of the photomultiplier, ALL your readings have to be performed using the same voltage. If one of your samples exceeds the scale of the fluorometer, setup the equipment again as I explained before, and read all your samples again. Don't try to use conversion factors, equivalences, or other mathematical manipulation of data. They just won't work.
I am using LS 45 and recently it started showing 'overflow' message before the scanning. I can not change slit value in LS 45, it is fixed at 10 nm. I tried using the minimum PMT voltage, but the problem persists. I tried with only distilled water, so no question of over concentration. Can anyone solve the problem.
One question: where the excitation and emission wavelengths too close? Did they overlap? I ask you this because if both wavelengths overlap the signal will be spurious and it will cause the "overflow" message.
You can try to check if your fluorometer is working OK using a very diluted alkaline solution of fluorescein (I don't recall now the exact concentration, but is in the nanomolar range).
Try running excitation and emission spectra of diluted fluorescein, adjusting the voltage of the photomultiplier. If you still have the error message, you have not choice but to ask PerkinElmer to send somebody to fix your fluorometer.
If fluorescein works fine, then the problem is the probe you are working with (which one is it, by the way? please, let me know because if I have some in my lab I can check it with my LS55).
I have a problem with a spectrofluorometer LS55 PerkinElmer.
The SCAN FLWINLAB application program does not run. And an error message is displayed: DDE Setup: Client failled to connect to server Lamp September. FL = On.
Yes, I have the same bug with a LS-55 Perkinelmer spectrofluorometer and the software FL Winlab v4.0. The bug is due to a software problem with the file names, if you use simple short names without dots, spaces, etc, to save your spectra you will avoid this bug.