I am synthesizing a nano composite. Will IR data give any quantitative information in regards of the filler concentration? Do peak area plays any role in contrast with the concentration?
Dera Varun, if the molar aborption coeffeicient does not change with concenration, you can get quantitative infomation but only if you prepare a calibration curve by producing samples of different, precisely known concentration. Don't forget that you have plot aborbance (ant not transmission) against the concentration.
Dera Varun, if the molar aborption coeffeicient does not change with concenration, you can get quantitative infomation but only if you prepare a calibration curve by producing samples of different, precisely known concentration. Don't forget that you have plot aborbance (ant not transmission) against the concentration.